MethyPrecise® Probe-Based Methylation Detection qPCR Master Mix (2× Methy-qPCR Mix) (JSR-E-206, JSR-E-207)

This product is the MethyPrecise® probe‑based methylation detection qPCR master mix (2× Methy‑qPCR Mix, catalog number JSR‑E‑206 series), specifically optimized for DNA methylation analysis. It is formulated with a genetically engineered hot‑start enzyme and a proprietary buffer, enhancing bisulfite‑template compatibility, GC‑content tolerance (25%–70%), uracil resistance, and 3′‑end mismatch discrimination, while delivering superior assay sensitivity and specificity.

Product item number

Packaging Specifications

Price

JSR-E-206

100 times

580

JSR-E-207

500 times

2680


Keywords



Product Description

FAQ

Product Information

Related Documents

Jianshi Bio MethyPrecise® Probe-Based Methylation Detection qPCR Premix (JSR-E-206 Series)

Product Overview

This product is the MethyPrecise® probe‑based methylation detection qPCR master mix (2× Methy‑qPCR Mix, catalog no. JSR‑E‑206 series), specifically optimized for DNA methylation analysis. It is formulated with a genetically engineered hot‑start enzyme and a proprietary buffer, enhancing bisulfite‑template compatibility, GC‑content tolerance (25%–70%), uracil resistance, and 3′‑end mismatch discrimination, while delivering exceptional sensitivity and specificity. The master mix is supplied as a ready‑to‑use 2× formulation; users need only add primers, probes, and template. It comes with two ROX reference dye options—high and low—and is compatible with mainstream qPCR instruments. Amplification curves exhibit excellent linearity, quantitative accuracy, and high reproducibility, making it suitable for assays such as methylation site detection and SNP genotyping. We recommend using this product in conjunction with the JSR‑530 Bisulfite Modification Kit. Available in 100‑ and 500‑reaction formats, store at −25°C to −15°C; intended solely for research use.

Product Features

  • Specifically optimized for methylation detection, this product combines a genetically engineered hot-start enzyme with a dedicated buffer to significantly enhance multiplex target‑specific detection performance in methylation‑based assays.
  • Compatible with a broad GC content range, it is suitable for templates with 25%–70% GC content, covering the methylation detection needs of diverse genomic regions.
  • Designed for robust tolerance, it exhibits excellent compatibility with sulfite conversion products and uracil‑modified templates, thereby minimizing template‑inhibition effects.
  • It exhibits strong recognition of mismatched bases at the 3′ end, enabling precise differentiation between methylated and unmethylated sites and achieving high detection specificity.
  • High sensitivity with low‑copy detection; excellent multiplex target‑detection sensitivity; even low‑abundance methylated templates can be stably detected.
  • Excellent amplification performance, with a highly linear amplification curve, accurate quantification of the target gene, and high reproducibility of experimental results.
  • 2× ready-to-use pre-mix; simply add primers, probes, and template to streamline the workflow and minimize pipetting errors.
  • Comes with dual-concentration ROX reference dye, available in high and low concentrations, and compatible with mainstream qPCR instruments from various brands—no additional purchase required.
  • The reaction system is highly flexible and customizable, with a 5–25 μL reaction volume that can be scaled proportionally to accommodate varying experimental throughput and sample‑size requirements.
  • Strict quality control ensures a one-year warranty; the product delivers stable and reliable performance and is intended solely for research and experimental use, not for clinical diagnostics.

Performance Verification Data

Using a methylation‑qPCR assay, amplification was performed at two genomic loci, A and B, with each template analyzed in duplicate. The results were compared with those obtained using a competing product of the same type, Y, to evaluate the assay’s amplification efficiency, linearity, and detection sensitivity.

Comparison of Amplification Performance of Gene A

Competitor Y of the same type

Jianshi JSR-E-206

Brand Conversion input quantity Cq 1 Cq 2 Average Cq (Avg)
Competitor Y of the same type 1ng 36.10 36.95 36.53
10ng 33.34 34.09 33.71
20ng 32.54 32.27 32.41
100ng 29.46 29.54 29.50
200ng 28.36 28.69 28.53
300ng 27.95 27.63 27.79
500ng 27.10 27.01 27.05
Jianshi JSR-E-206 1ng 36.02 39.18 37.60
10ng 32.72 32.41 32.56
20ng 31.85 32.02 31.94
100ng 28.91 28.76 28.83
200ng 27.90 27.86 27.88
300ng 27.17 27.37 27.27
500ng 26.31 26.36 26.34

Comparison of B Gene Amplification Performance

Competitor Y of the same type

Jianshi JSR-E-206

Brand Conversion input quantity Cq 1 Cq 2 Average Cq (Avg)
Competitor Y of the same type 1ng 37.09 35.68 36.39
10ng 32.73 33.28 33.01
20ng 31.53 31.31 31.42
100ng 28.69 28.85 28.77
200ng 27.89 27.70 27.79
300ng 27.01 26.91 26.96
500ng 26.52 26.17 26.35
Jianshi JSR-E-206 1ng 36.32 36.39 36.36
10ng 32.38 31.62 32.00
20ng 30.74 30.89 30.81
100ng 27.81 27.74 27.77
200ng 26.61 26.85 26.73
300ng 26.11 25.30 25.71
500ng 25.30 25.26 25.28

Jianshi Bio MethyPrecise® Probe-Based Methylation Detection qPCR Premix

2×Methy-qPCR Mix: Specialized Reagent for Methylation Quantitative Detection – Frequently Asked Questions

Core Performance Features

  • Methylation-specific : Hot-start enzyme + optimized buffer, suitable for methylation detection assays
  • Highly tolerant and widely compatible : Sulfite-resistant, compatible with templates containing 25–70% GC content
  • High sensitivity and high accuracy : Strong capability for detecting low-copy targets, excellent amplification linearity, and precise quantification.
  • Easy to operate : 2× pre-mix system; only primers, probes, and template need to be added.

Item Number, Specifications, and Composition

  • JSR-E-206 : 100 reactions, 1 mL pre-mix
  • JSR-E-207 : 500 reactions, 5 × 1 mL aliquots of premix
  • Matching dye : Comes with two reference dyes, 50× ROX High and Low.
  • Storage conditions : Store frozen and protected from light at -25 to -15°C.

Applicable experimental scenarios

  • Methylation detection : Quantitative detection of DNA methylation sites
  • Genotyping : SNP Detection and Allele Typing
  • Conversion sample : Sulfite-converted DNA sample analysis
  • Low-copy detection : Sensitive detection of trace methylated templates

High-Frequency FAQs for Methylation qPCR Master Mix

1: What is this MethyPrecise® qPCR Master Mix, and what are its key advantages?

This product is a 2× probe-based qPCR master mix specifically optimized for DNA methylation detection. It employs an engineered hot-start enzyme and a proprietary buffer system, significantly enhancing the assay performance for methylated samples.

1. Formulation Optimization: Composed of a genetically engineered hot-start enzyme and an optimized reaction buffer, specifically tailored for methylation detection applications.

2. Tolerance and compatibility: Exhibits strong sulfite tolerance, is compatible with templates spanning 25–70% GC content, and demonstrates robust uracil tolerance as well as the ability to recognize 3′-terminal mismatched bases.

3. Excellent performance: Multi‑target sensitivity and specificity, as well as the ability to detect low‑copy‑number targets, have been significantly enhanced; amplification curves exhibit outstanding linearity, and quantitative accuracy for target genes is high.

4. Convenient operation: A 2× pre-mix system requires only the addition of primers, probes, and template during use, significantly reducing pipetting steps and minimizing experimental errors.

2: What are the product’s item numbers and specifications? What core components are included in the kit?

Two reaction‑scale options are available: a pre‑mix plus a dual‑concentration ROX reference dye kit, eliminating the need to purchase additional reagents.

1. Item number and specifications:
• JSR-E-206: 100 reactions specification, including 1 mL of 2× pre-mix;
• JSR-E-207: 500 reactions, including 5 × 1 mL of 2× pre-mix, for convenient aliquoting and use;

2. Core components:
• 2×Methy-qPCR Mix: Pre-mixed with hot-start DNA polymerase, dNTPs, an optimized buffer system, and more;
• 50×ROX High: High-concentration reference dye;
• 50×ROX Low: Low-concentration reference dye.

Supporting Tips : Two types of ROX are included with the kit; select the appropriate one based on your instrument model.
3: What are the storage conditions for this product? What precautions should be taken before use?

Standardized storage and handling are critical to ensuring enzyme activity and amplification stability.

1. Storage conditions: Store frozen at -25°C to -15°C, protected from light; a one-year quality guarantee is provided from the date of sale.

2. Pre-use preparation: After removal, place on ice to thaw; once completely thawed, gently invert to mix, then briefly centrifuge before use.

3. Instrument‑use requirements: Prepare the reaction mixture, securely cap the tubes, and briefly centrifuge them to remove air bubbles that could interfere with fluorescence signal acquisition.

4. Post-use handling: Immediately replace the cap and store at −20°C to prevent prolonged exposure of the reagent to room temperature.

Important Reminder : Avoid repeated freeze-thaw cycles, as they can reduce enzyme activity; it is recommended to aliquot the product into single-use portions for storage.
4: How should a standard qPCR reaction mix be prepared? Can the reaction mix be optimized or adjusted?

A standard 20 μL reaction system with clearly defined component ratios, which can be adjusted as needed within a specified range.

1. Standard 20 μL reaction system:
• 2×Methy-qPCR Mix: 10 μL
• Forward Primer (10 μM): 0.4 μL
• Reverse Primer (10 μM): 0.4 μL
• Probe (10 μM): 0.4 μL
• Template DNA: Add as needed (recommended 2 μL)
• 50×ROX (Low/High): 0.4 μL
• ddH₂O: bring to 20 μL

2. System Adjustment: The reaction volume can be scaled proportionally within the range of 5–25 μL; volumes exceeding 50 μL are not recommended.

3. Concentration Optimization: The final concentrations of primers and probes can be adjusted within the range of 0.1–1 μM to achieve optimal amplification performance.

Operational Recommendations : Pre-mix the common components first, then aliquot them to minimize pipetting errors and improve assay reproducibility.
5: How should you choose between high‑ and low‑ROX reference dyes? Which instruments are each compatible with?

ROX is used for fluorescence signal baseline correction and must be selected to match the specific qPCR instrument model to ensure quantitative accuracy.

1. ROX High : Compatible with Applied Biosystems® models 5700, 7000, 7300, 7700, 7900HT, StepOne™, StepOnePlus™, and others;

2. ROX Low : Compatible with Applied Biosystems® 7500, ViiA™ 7, QuantStudio™ 12K Flex, Agilent Mx3000P™, Mx3005P™, Mx4000™, and other models;

3. No ROX required : Compatible with Rotor-Gene™, Opticon™, Chromo4™, LightCycler® series, Bio-Rad CFX96, Illumina Eco™, and other models.

Attention : An incorrect ROX selection can lead to baseline abnormalities and non-smooth curves, severely compromising the accuracy of quantitative results.
6: What are the requirements for the template when using this pre-mix? What is the compatible GC content range?

Optimized for methylated samples, with broad GC tolerance and high bisulfite resistance, it is compatible with a wide range of post‑conversion DNA samples.

1. Applicable templates: DNA samples after sulfite conversion, methylated standards, plasmid templates, genomic DNA, etc.;

2. GC compatibility range: Compatible with templates having 25% to 70% GC content; targets with varying GC contents can all be stably amplified.

3. Tolerance characteristics: Exhibits excellent uracil tolerance, strong recognition of 3′-terminal mismatched bases, and is well suited to the base‑specific properties of DNA following bisulfite conversion.

4. Template recommendation: For methylation analysis, we recommend using our company’s JSR530 kit for bisulfite conversion.

7: What is the standard qPCR reaction protocol, and how can its parameters be optimized?

The standard two-step amplification protocol allows for flexible adjustment of parameters based on the template’s GC content and the primers’ Tm values.

1. Standard qPCR protocol (two-step method):
• Pre-denaturation: 95°C for 30 seconds, 1 cycle;
• Amplification cycles: 95°C denaturation for 10 seconds, followed by 60°C annealing/extension for 30 seconds with fluorescence acquisition, for a total of 35–45 cycles.

2. Parameter Optimization:
• Denaturation: For high-GC templates, the denaturation time can be extended to 3–5 minutes to ensure complete template denaturation.
• Annealing temperature: Adjust according to the primer/probe Tm value, with a range of 50–65°C; for initial experiments, a preliminary run at 60°C is recommended.
• Number of cycles: Typically 35–40 cycles; for low-copy-number samples, the cycle number may be increased to 45.

Collection Settings : The two-step method collects signals during the annealing/extension phase, while the three-step method collects signals during the extension phase.
8: What are the causes of an uneven amplification curve, a sudden drop or decline in the curve, and an abrupt decrease at certain points? How can these issues be resolved?

Abnormal amplification curves are often caused by issues such as instrument calibration, template concentration, air bubbles, or baseline settings; these can be systematically investigated and resolved.

1. The curve is not smooth. The reasons are system uncalibration, template concentration too low, and incorrect ROX selection.
Solution: Recalibrate the instrument’s fluorescence system, increase the template concentration, and verify that ROX is compatible with the instrument.

2. Curve breakage/decline The reasons are excessively high template concentration and an inappropriate baseline selection range.
Solution: Reduce the template concentration, reset the baseline range, and avoid regions with fluorescence fluctuations.

3. A sudden drop in individual curves The reason is that bubbles in the reaction tube interfere with fluorescence detection.
Solution: Before loading the sample, briefly centrifuge the PCR tube to remove air bubbles from the bottom and walls.

9: What are the causes of no amplification curve and elevated Ct values at the end of the reaction, and how can these issues be troubleshooted?

No amplification or elevated Ct values can stem from various factors, including the template, reagents, reaction conditions, and primers; systematically troubleshooting each step will help quickly pinpoint the issue.

1. No amplification curve Common causes and solutions:
• Insufficient cycle number: It is recommended to set ≥40 cycles;
• Signal acquisition not configured: Verify that fluorescence detection is enabled during the annealing‑extension step of the two‑step protocol.
• Primer degradation: Prимер integrity is assessed by PAGE electrophoresis; if degradation is detected, the primer is resynthesized.
• Template concentration too low/degradation: Increase the template loading amount, or prepare the template anew.

2. High Ct value Common causes and solutions:
• Low amplification efficiency: Optimize reaction conditions, try a three-step amplification protocol, or redesign the primers.
• Low template concentration: reduce template dilution and increase the amount loaded.
• Overly long PCR products: It is recommended to keep the amplicon length between 80 and 150 bp.
• Presence of PCR inhibitors: Increase the template dilution factor, or re-purify the template.

10: How can the issues of negative control amplification, poor linearity of the standard curve, and poor experimental reproducibility be addressed, respectively?

Three common types of experimental issues correspond to different optimization strategies:

1. A clear amplification was observed in the negative control. :
Cause: Contamination of the reaction system and non-specific primer design.
Solution: Replace the reagents, water, and primers; perform all procedures within a clean bench to minimize aerosol contamination; and redesign the primers to enhance specificity.

2. The standard curve exhibits poor linearity. :
Causes: pipetting errors, degradation of standards, and excessively high template concentrations.
Solution: Increase the template dilution factor and the sample volume, re‑prepare the standards, and optimize the gradient concentration range.

3. The experiment has poor reproducibility. :
Reasons: inconsistent sample volumes and excessively low template concentrations.
Solution: Use a pipette with higher precision, reduce the template dilution factor, or increase the aliquot volume; pre‑mix common reagents and then aliquot.

Related Products


Get a Quote


If you are interested in our products, please leave your email address, and we will contact you as soon as possible. Thank you!

Submit
Image Name

Order Consultation: Hu Manager

Image Name
Image Name

Order Consultation: An Manager

Image Name
Image Name

End-user Scientific Research Customer Order Consultation

Image Name
Image Name

Technical Consultation: Bai Technical

Image Name
Image Name

Customer Service

Image Name
Image Name

Douyin

Image Name
Image Name

WeChat Video Account

Image Name
Image Name

Kuaishou

Image Name

Contact Us

Jianshi Biotechnology (Beijing) Co., Ltd.
Company Address: Room 401, Building 11, No. 12 Courtyard, Juyuan Middle Road, Mapo Town, Shunyi District, Beijing  
Phone: 86 18640171737
E-mail: gwang@jianshibio.com
Jianshi Biotechnology (Zhejiang) Co., Ltd.
Address: Building 11, No. 36 Xingshun Road, Zhongyi Ningbo Eco-park, Yuyao City, Zhejiang Province 
WhatsApp: 86 18640171737

Leave a Message

If you are interested in our products, please leave your phone number.

Consult Now →

Product Subscription

Subscribe to us to stay informed about product updates and special offers

%{tishi_zhanwei}%

Copyright: Jianshi Biotechnology (Beijing) Co., Ltd.    Powered by:CEglobal

Business License

WhatsApp:

mzhao@jianshibio.com:

WeChat:

8601058235289:

Product Inquiry

Filling in your phone number and email address will help us contact you promptly and resolve your questions as quickly as possible.